Characterization of microMOSFET detectors for in vivo dosimetry in high-dose-rate brachytherapy with 192 Ir

The objective of this study was to characterize the Best Medical Canada microMOSFET detectors for their application in in vivo dosimetry for high-dose-rate brachytherapy (HDRBT) with Ir. We also developed a mathematical model to correct dependencies under the measurement conditions of these detector...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2020-06, Vol.47 (5), p.2242-2253
Hauptverfasser: Ruiz-Arrebola, Samuel, Fabregat-Borrás, Rosa, Rodríguez, Eduardo, Fernández-Montes, Manuel, Pérez-Macho, Mercedes, Ferri, María, García, Ana, Cardenal, Juan, Pacheco, María T, Anchuelo, Javier, Tornero-López, Ana M, Prada, Pedro J, Guirado, Damián
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The objective of this study was to characterize the Best Medical Canada microMOSFET detectors for their application in in vivo dosimetry for high-dose-rate brachytherapy (HDRBT) with Ir. We also developed a mathematical model to correct dependencies under the measurement conditions of these detectors. We analyzed the linearity, reproducibility, and interdetector variability and studied the microMOSFET response dependence on temperature, source-detector distance, and angular orientation of the receptor with respect to the source. The correction model was applied to 19 measurements corresponding to five simulated treatments in a custom phantom specifically designed for this purpose. The detectors (high bias applied in all measurements) showed excellent linearity up to 160 Gy. The response dependence on source-detector distance varied by (8.65 ± 0.06)% (k = 1) for distances between 1 and 7 cm, and the variation with temperature was (2.24 ± 0.05)% (k = 1) between 294 and 310 K. The response difference due to angular dependence can reach (10.3 ± 1.3)% (k = 1). For the set of measurements analyzed, regarding angular dependences, the mean difference between administered and measured doses was -4.17% (standard deviation of 3.4%); after application of the proposed correction model, the mean difference was -0.1% (standard deviation of 2.2%). For the treatments analyzed, the average difference between calculations and measures was 4.7% when only the calibration coefficient was used, but it is reduced to 0.9% when the correction model is applied. Important response dependencies of microMOSFET detectors used for in vivo dosimetry in HDRBT treatments, especially the angular dependence, can be adequately characterized by a correction model that increases the accuracy of this system in clinical applications.
ISSN:0094-2405
2473-4209
DOI:10.1002/mp.14080