The accuracy of bone mineral density measurement using dual-energy spectral CT and quantitative CT: a comparative phantom study
To compare bone mineral density (BMD) measurement accuracy by dual-energy spectral computed tomography (CT) and quantitative CT (QCT) using an anthropomorphic European spine phantom (ESP). An ESP containing three hydroxyapatite inserts with densities of 50, 100, and 200 mg/cm3 was scanned 10 times e...
Gespeichert in:
Veröffentlicht in: | Clinical radiology 2020-04, Vol.75 (4), p.320.e9-320.e15 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To compare bone mineral density (BMD) measurement accuracy by dual-energy spectral computed tomography (CT) and quantitative CT (QCT) using an anthropomorphic European spine phantom (ESP).
An ESP containing three hydroxyapatite inserts with densities of 50, 100, and 200 mg/cm3 was scanned 10 times each using spectral CT and QCT protocols. Their BMD values were measured using hydroxyapatite-based spectral CT material decomposition images and QCT images and compared with the true values in ESP.
Both protocols had good repeatability in BMD measurement with low coefficient-of-variance (spectral CT: |
---|---|