In Vivo Evaluation of 3 Electronic Apex Locators: Root ZX Mini, Apex ID, and Propex Pixi
The aim of this in vivo study was to compare the accuracy of 3 electronic apex locators (EALs) (Root ZX mini [J Morita Corp, Tokyo, Japan], Apex ID [SybronEndo, Glendora, CA], and Propex Pixi [Dentsply Maillefer, Ballaigues, Switzerland]) to determine the working length. Thirty single-rooted human t...
Gespeichert in:
Veröffentlicht in: | Journal of endodontics 2020-02, Vol.46 (2), p.158-161 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this in vivo study was to compare the accuracy of 3 electronic apex locators (EALs) (Root ZX mini [J Morita Corp, Tokyo, Japan], Apex ID [SybronEndo, Glendora, CA], and Propex Pixi [Dentsply Maillefer, Ballaigues, Switzerland]) to determine the working length.
Thirty single-rooted human teeth that were scheduled for extraction were selected for the study. Electronic measurements were performed with the 3 EALs. After the teeth had been extracted, a #10 K-file was used to determine the actual working length, which was established at 0.5 mm short of the major foramen. The data were statistically analyzed with analysis of variance (α = 0.05).
No significant differences were found among the experimental groups (P > .05). The mean distance from the actual working length to the file tip was 0.163 ± 0.032 mm when Root ZX mini was used, 0.343 ± 0.032 mm for Propex Pixi , and 0.012 ± 0.008 mm for Apex ID.
Under the in vivo conditions of this study, no statistically significant differences were observed among the 3 EALs. |
---|---|
ISSN: | 0099-2399 1878-3554 |
DOI: | 10.1016/j.joen.2019.10.035 |