A further development of the fluorescent gated gas scintillation proportional counter
Recent results from a K gated gas scintillation proportional counter filled with a xenon/helium mixture to a pressure of about 3 bar have already shown that a background rejection efficiency in excess of 99.3% can be achieved in the energy range 35-100 keV. In addition the application of the techniq...
Gespeichert in:
Veröffentlicht in: | IEEE Trans. Nucl. Sci.; (United States) 1983-02, Vol.NS-30, p.394-397 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Recent results from a K gated gas scintillation proportional counter filled with a xenon/helium mixture to a pressure of about 3 bar have already shown that a background rejection efficiency in excess of 99.3% can be achieved in the energy range 35-100 keV. In addition the application of the technique of fluorescent gating also provides energy resolutions below 3% at about 60 keV. Experimental results are presented for a gas scintillation counter filled with a krypton/xenon mixture which, through the use of the krypton K fluorescent gate, enables a high background rejection and improved energy resolution, for the energy range above 14 keV, to be achieved. |
---|---|
ISSN: | 0018-9499 |
DOI: | 10.1109/TNS.1983.4332297 |