Design of a Real-time Self-adjusting Calibration Algorithm to Improve the Accuracy of Continuous Blood Glucose Monitoring
The aim of this study is to establish a real-time self-adjusting calibration algorithm to compensate for signal drift and sensitivity attenuation of subcutaneous implantable glucose sensors. A real-time self-adjusting in vivo calibration method was designed based on the one-point calibration model....
Gespeichert in:
Veröffentlicht in: | Applied biochemistry and biotechnology 2020-04, Vol.190 (4), p.1163-1176 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this study is to establish a real-time self-adjusting calibration algorithm to compensate for signal drift and sensitivity attenuation of subcutaneous implantable glucose sensors. A real-time self-adjusting in vivo calibration method was designed based on the one-point calibration model. The current signal was compensated in real-time and the sensitivity was calibrated regularly. The least squares method was used to fit the initial parameters of the model, and then, the in vivo monitored current data was calibrated. Comparing with the mean absolute relative difference (MARD) of the blood glucose concentration by the traditional one-point calibration model (22.85 ± 5.76%), the MARD of the blood glucose concentration calibrated by the real-time self-adjusting in vivo calibration method was 6.28 ± 2.31%. The accuracy of the dynamic blood glucose monitoring was effectively improved. This calibration algorithm could compensate the signal drift in real time and correct sensitivity regularly to improve the accuracy of dynamic glucose monitoring, thus significantly enhancing diabetic self-management. |
---|---|
ISSN: | 0273-2289 1559-0291 |
DOI: | 10.1007/s12010-019-03142-7 |