Contrast-enhanced modified 3D T1-weighted TSE black-blood imaging can improve detection of infectious and neoplastic meningitis
Objectives To evaluate the diagnostic value of a contrast-enhanced 3D T1-weighted-modified volumetric isotropic turbo spin-echo acquisition sequence (T1-mVISTA) in comparison with a conventional 3D T1-weighted magnetization-prepared rapid gradient-echo (T1-MP-RAGE) sequence for the detection of meni...
Gespeichert in:
Veröffentlicht in: | European radiology 2020-02, Vol.30 (2), p.866-876 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objectives
To evaluate the diagnostic value of a contrast-enhanced 3D T1-weighted-modified volumetric isotropic turbo spin-echo acquisition sequence (T1-mVISTA) in comparison with a conventional 3D T1-weighted magnetization-prepared rapid gradient-echo (T1-MP-RAGE) sequence for the detection of meningeal enhancement in patients with meningitis.
Methods
Thirty patients (infectious meningitis,
n
= 12; neoplastic meningitis,
n
= 18) and 45 matched controls were enrolled in this retrospective case-control study. Sets of randomly selected T1-mVISTA and T1-MP-RAGE images (both with 0.8-mm isotropic resolution) were read separately 4 weeks apart. Image quality, leptomeningeal and dural enhancement, grading of visual contrast enhancement, and diagnostic confidence were compared using the Kruskal-Wallis rank sum test.
Results
Image quality was rated to be good to excellent in 75 out of 75 cases (100%) for T1-mVISTA and 74 out of 75 cases (98.7%) for T1-MP-RAGE. T1-mVISTA detected significantly more patients with leptomeningeal enhancement (
p
= 0.006) compared with T1-MP-RAGE (86.7 vs. 50.0%,
p
|
---|---|
ISSN: | 0938-7994 1432-1084 |
DOI: | 10.1007/s00330-019-06475-3 |