Clinical study of the novel FlexitSystem implant for high tibial open wedge osteotomy

The FlexitSystem implant is a novel implant used in open wedge high tibial osteotomy. A clinical safety study was performed. Retrospectively 50 patients were analyzed who were treated with an open wedge high tibial osteotomy and the new FlexitSystem implant, with a minimal follow-up of one year. Com...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acta orthopaedica belgica 2019-09, Vol.85 (3), p.381-386
Hauptverfasser: Van Egmond, Nienke, Van De Groes, Sebastiaan Aw, Van Kampen, Albert
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The FlexitSystem implant is a novel implant used in open wedge high tibial osteotomy. A clinical safety study was performed. Retrospectively 50 patients were analyzed who were treated with an open wedge high tibial osteotomy and the new FlexitSystem implant, with a minimal follow-up of one year. Complication rate, radiographic outcomes and implant removal were investigated. One patient underwent a revision surgery because of loss of correction and non-union. The complication rate was 10.0%. No other radiographic complications (screw breakage, implant failure) were found. In 24 patients (48%) the FlexitSystem implant was removed at a mean follow-up of 12.6 months (range 2.6 till 24.0 months). The mean reason was irritation of the implant. The FlexitSystem implant is a clinical safe and stable implant for an open wedge high tibial osteotomy, with a low complication rate. The rate of implant irritation requiring removal remained high.
ISSN:0001-6462