Correction of head movement by frame-to-frame image realignment for receptor imaging in positron emission tomography studies with [11C]raclopride and [11C]FLB 457
Objective Positron emission tomography (PET) scans of imaging receptors require 60–90-min dynamic acquisition for quantitative analysis. Head movement is often observed during scanning, which hampers the reliable estimation of quantitative parameters. This study evaluated image-based motion correcti...
Gespeichert in:
Veröffentlicht in: | Annals of nuclear medicine 2019-12, Vol.33 (12), p.916-929 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
Positron emission tomography (PET) scans of imaging receptors require 60–90-min dynamic acquisition for quantitative analysis. Head movement is often observed during scanning, which hampers the reliable estimation of quantitative parameters. This study evaluated image-based motion correction by frame-to-frame realignment for PET studies with [
11
C]raclopride and [
11
C]FLB 457 acquired by an Eminence SET-3000GCT/X and investigated the effect of this correction on the quantitative outcomes.
Methods
First, an optimal method for estimating motion parameters was evaluated by computer simulation. Simulated emission sinograms were reconstructed to the PET images with or without attenuation correction using a µ-map of the transmission scan. Six motion parameters were estimated frame-by-frame by registering each frame of the PET images to several types of reference images and the reliability of registration was compared. Next, in [
11
C]raclopride and [
11
C]FLB 457 studies in normal volunteers, six motion parameters for each frame were estimated by the registration method determined from the simulation results. Head movement was corrected by realigning the PET images reconstructed with a motion-included µ-map in which a mismatch between the transmission and emission scans was corrected. After this correction, time-activity curves (TAC) for the striatum or cerebral cortex were obtained and the binding potentials of the receptors (BP
ND
) were estimated using the simplified reference tissue model.
Results
In the simulations, the motion parameters could be reliably estimated by registering each frame of the non-attenuation-corrected PET images to their early-phase frame. The motion parameters in the human studies were also obtained using the same method. After correction, a discontinuity of TACs in the striatum and cerebral cortex was remarkably improved and the BP
ND
values in these regions increased. Compared to the motion-corrected PET images reconstructed using the measured µ-map, the images reconstructed using the motion-included µ-map did not result in a remarkable improvement of BP
ND
in the striatum of [
11
C]raclopride studies, while the BP
ND
in the cerebral cortex changed in some [
11
C]FLB 457 studies in which large head movement was observed.
Conclusions
In PET receptor imaging, head movement during dynamic scans can be corrected by frame-to-frame realignment. This method is easily applicable to clinical studies and provides reliable TACs and |
---|---|
ISSN: | 0914-7187 1864-6433 |
DOI: | 10.1007/s12149-019-01405-1 |