Microdosimetry at the CATANA 62 MeV proton beam with a sealed miniaturized TEPC
•Miniature Tissue Equivalent Proportional Counters can work without gas-flow.•The new sealed mini-TEPC gives reproducible results over a period of 4 months.•The new mini-TEPC can be used as a precise LET-monitor in proton therapy.•The new mini-TEPC measurements result in a reasonable RBE-assessment....
Gespeichert in:
Veröffentlicht in: | Physica medica 2019-08, Vol.64, p.114-122 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Miniature Tissue Equivalent Proportional Counters can work without gas-flow.•The new sealed mini-TEPC gives reproducible results over a period of 4 months.•The new mini-TEPC can be used as a precise LET-monitor in proton therapy.•The new mini-TEPC measurements result in a reasonable RBE-assessment.•The mini-TEPC can be used for a comparative study of different radiotherapy centres.
A new mini-TEPC with cylindrical sensitive volume of 0.9 mm in diameter and height, and with external diameter of 2.7 mm, has been developed to work without gas flow. With such a mini counter we have measured the physical quality of the 62 MeV therapeutic proton beam of CATANA (Catania, Italy). Measurements were performed at six precise positions along the Spread-Out Bragg Peak (SOBP): 1.4, 19.4, 24.6, 29.0, 29.7 and 30.8 mm, corresponding to positions of clinical relevance (entrance, proximal, central, and distal-edge of the SOBP) or of high lineal energy transfer (LET) increment (distal-dose drop off). Without refilling the microdosimeter with new gas, the measurements were repeated at the same positions 4 months later, in order to study the stability of the response in sealed-mode operation.
From the microdosimetric spectra the frequency-mean lineal energy y-F and the dose-mean lineal energy y-D were derived and compared with average LET values calculated by means of Geant4 simulations. The comparison points out, in particular, a good agreement between microdosimetric y-D and the total dose-average LET¯d, which is the average LET of the mixed radiation field, including the contribution by nuclear reactions. |
---|---|
ISSN: | 1120-1797 1724-191X |
DOI: | 10.1016/j.ejmp.2019.06.011 |