Three-dimensional high-resolution simultaneous quantitative mapping of the whole brain with 3D-QALAS: An accuracy and repeatability study
Previous methods for the quantification of brain tissue properties by magnetic resonance imaging were mainly based on two-dimensional acquisitions and were thus limited to a relatively low resolution in the slice direction compared to three-dimensional (3D) acquisitions. The 3D-quantification using...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance imaging 2019-11, Vol.63, p.235-243 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Previous methods for the quantification of brain tissue properties by magnetic resonance imaging were mainly based on two-dimensional acquisitions and were thus limited to a relatively low resolution in the slice direction compared to three-dimensional (3D) acquisitions. The 3D-quantification using an interleaved Look–Locker acquisition sequence with a T2 preparation pulse (3D-QALAS) sequence may allow for simultaneous acquisition of relaxometry parameters in high spatial resolution.
To evaluate bias, linearity, and day-to-day repeatability of relaxometry parameters, as well as tissue fraction maps, acquired with 3D-QALAS.
Scan–rescan test of the 3D-QALAS sequence was performed on a 1.5-T scanner with the International Society for Magnetic Resonance in Medicine/National institute of Standards and Technology system phantom and 10 healthy volunteers (7 male, 3 female; mean age, 23.2 ± 3.6 years). Simple linear regression analysis, Bland–Altman plots, and intrasubject coefficients of variation (CV) were used to assess the reliability of 3D-QALAS sequence-derived parameters. The T1, T2, proton density (PD), and myelin volume fraction (MVF) of in vivo brain regions were compared with values obtained using the multidynamic multi-echo sequence.
In the phantom study, the T1, T2, and PD values measured by 3D-QALAS showed strong linearity with the reference values (R2 = 0.998, 0.998, and 0.960 for T1, T2, and PD, respectively) and high repeatability (mean CV of 1.2%, 2.8%, and 2.9% for T1, T2, and PD, respectively). The T1, T2, PD, and MVF values of in vivo brain regions obtained with 3D-QALAS were highly consistent within subjects, with mean intrasubject CVs of 0.5%, 0.5%, 0.4%, and 1.6% for the T1, T2, PD, and MVF values, respectively.
3D-QALAS enables reliable measurement of T1, T2, PD, and MVF values of the whole brain in high spatial resolution across a clinically-relevant dynamic range. |
---|---|
ISSN: | 0730-725X 1873-5894 |
DOI: | 10.1016/j.mri.2019.08.031 |