Resolution limits in Fourier Transform Spectrometry
It is shown by mathematical analysis, and by close inspection of experimentally determined spectra, that the frequently quoted result Δ v̄= 0.5/ D, where D is the maximum path difference introduced, for the resolution in unapodized Fourier transform spectrometry is over optimistic. Even when the two...
Gespeichert in:
Veröffentlicht in: | Infrared physics 1976-01, Vol.16 (6), p.655-660 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | It is shown by mathematical analysis, and by close inspection of experimentally determined spectra, that the frequently quoted result Δ
v̄= 0.5/
D, where
D is the maximum path difference introduced, for the resolution in unapodized Fourier transform spectrometry is over optimistic. Even when the two lines are of equal strength the resolution is found to be Δ
v̄ =0.7/
D whilst if they differ markedly in strength the resolution is even worse (∼ 0.8/
D). The loss in resolution in going to apodized spectrometry is not as severe as has previously been thought with Δ
v̄ = 0.9/
D for equal strength components and Δ
v̄ = 1/
D for the components having an intensity ratio of 2:1. |
---|---|
ISSN: | 0020-0891 |
DOI: | 10.1016/0020-0891(76)90102-0 |