How to CT scan human mummies: Theoretical considerations and examples of use
The purpose of this study was to develop and provide recommendations for computed tomography (CT) examinations of human mummies. Based on theoretical considerations and examples, recommendations for use are provided for scanning and image reconstruction parameters. Four examples are presented to ill...
Gespeichert in:
Veröffentlicht in: | International journal of paleopathology 2019-09, Vol.26, p.122-134 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to develop and provide recommendations for computed tomography (CT) examinations of human mummies.
Based on theoretical considerations and examples, recommendations for use are provided for scanning and image reconstruction parameters. Four examples are presented to illustrate the effect of different CT parameters on image quality.
The use of appropriate scanning parameters (detector collimation, pitch factor, rotation time) is known to improve image quality; technical considerations favor using lower tube voltage and higher tube current values for the purposes of scanning of human mummies. The use of appropriate image reconstruction parameters (slice thickness, increment, field of view, reconstruction filters) are the basis for individual reconstructions for the purpose of evaluation, documentation, illustration and data storage.
Downsizing the field of view to the region of interest as done in the clinical radiological routine represents one major tool to improve image quality.
The provided recommendations should improve CT image quality in mummy studies as well as the handling of image data for reconstructions and storage.
The recommendations for CT scanning parameters and image reconstructions were written with relatively new generation CT scanners in mind. Only a few examples of use were chosen and image quality assessment was performed subjectively and not by quantitative measurements. |
---|---|