How comparable are total human chorionic gonadotropin (hCGt) tumour markers assays?

Background Total human chorionic gonadotropin (hCGt) tumour marker testing is regarded as an "off label" application for most commercial methods. We compared four assays in patients with a hCGt tumour marker request. We hypothesised that regression slopes would be altered and that outliers...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical chemistry and laboratory medicine 2020-03, Vol.58 (3), p.438-444
Hauptverfasser: Pretorius, Carel J., du Toit, Stephen, Wilgen, Urs, Klingberg, Sandra, Jones, Mark, Ungerer, Jacobus P.J., Tate, Jillian R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Total human chorionic gonadotropin (hCGt) tumour marker testing is regarded as an "off label" application for most commercial methods. We compared four assays in patients with a hCGt tumour marker request. We hypothesised that regression slopes would be altered and that outliers would be more common with tumour marker than with pregnancy samples if the detection of malignancy associated hCG molecular forms differed amongst assays. Further such systematic differences would be obvious and large enough to change clinical management decisions. Results We measured hCGt in 390 samples from 137 females and 253 males with a tumour marker request and 208 pregnancy controls with the following methods: Access Total βhCG, Architect Total-βhCG, Cobas hCG + β and Immulite HCG. The between method regressions determined on tumour marker and pregnancy samples were not significantly different. The outlier rates were similar for male and female tumour marker and the pregnancy groups: 1.6% (95% confidence interval [CI] 0%-3.1%), 2.2% (95% CI 0%-4.7%) and 2.9% (95% CI 0.6%-5.2%). The outliers were randomly distributed amongst the methods and we were confident that they would not adversely influence clinical decisions. Conclusions The hCGt results were clinically equivalent with no systematic difference amongst the four assays.
ISSN:1434-6621
1437-4331