Standardisation of 85Sr with digital anticoincidence counting and half-life determination of the 514 keV level of 85Rb
The activity of a 85Sr solution was determined by means of fully digital 4πβ(LS)-γ anticoincidence counting. The measurements were carried out in a custom-built, combined TDCR / 4πβ(LS)-γ coincidence system, utilising a commercially available CAEN N6751C digitizer. The analysis of the experimental d...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2019-11, Vol.153, p.108799-108799, Article 108799 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The activity of a 85Sr solution was determined by means of fully digital 4πβ(LS)-γ anticoincidence counting. The measurements were carried out in a custom-built, combined TDCR / 4πβ(LS)-γ coincidence system, utilising a commercially available CAEN N6751C digitizer. The analysis of the experimental data, collected in list-mode format, was performed off-line by using the SoftKAM computer code developed at PTB. The data were also used to determine the half-life of the 514 keV level of 85Rb which was found to be (1020.2 ± 6.0) ns.
•The activity of 85Sr was measured using 4πβ(LS)-γ anticoincidence counting.•The half-life of the 514 keV level of 85Rb was determined.•Measurements were carried out with a digitizer-based DAQ. |
---|---|