Oversensing issues leading to device extraction: When subcutaneous implantable cardioverter-defibrillator reached a dead-end
Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantations are rapidly expanding. However, the subcutaneous detection and interpretation of cardiac signals in S-ICDs is much more challenging than by conventional devices. There is a complete paradigm shift in cardiac signal sensing wit...
Gespeichert in:
Veröffentlicht in: | Heart rhythm 2020-01, Vol.17 (1), p.66-74 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantations are rapidly expanding. However, the subcutaneous detection and interpretation of cardiac signals in S-ICDs is much more challenging than by conventional devices. There is a complete paradigm shift in cardiac signal sensing with subcutaneous signal detection, leading in some cases to oversensing with restricted programming options.
The aim of this single-center study was to quantify and describe cases where recurring oversensing made the extraction of the device necessary.
Consecutive patients (n = 108) implanted with an S-ICD in our tertiary referral hospital were considered for analysis. Clinical and remote monitoring data were analyzed.
The S-ICD had to be explanted in 6 of 108 implanted patients (5.6%) because of refractory oversensing issues: myopotential oversensing, P- or T-wave oversensing, rate-dependent left bundle branch block aberrancy during exercise with R-wave double counting, and R-wave amplitude decrease after ventricular tachycardia ablation leading to noise detection. Seventeen of 108 patients experienced oversensing (15.7%): 9 patients had at least 1 inappropriate charge without a shock (8.3%), 3 patients had at least 1 inappropriate shock (2.8%), and 5 patients had both episodes (4.6%).
So far, cardiologists have had to deal with transvenous ICD lead fractures, but signal oversensing without correcting programming option could be the equivalent weakness of S-ICDs, despite an adequate screening.
[Display omitted] |
---|---|