Insufficient slow-flow suppression mimicking aneurysm wall enhancement in magnetic resonance vessel wall imaging: a phantom study
MR vessel wall imaging (VWI) is increasingly performed in clinical settings to support treatment decision-making regarding intracranial aneurysms. Aneurysm wall enhancement after contrast agent injection is expected to be related to aneurysm instability and rupture status. However, the authors hypot...
Gespeichert in:
Veröffentlicht in: | Neurosurgical focus 2019-07, Vol.47 (1), p.E19-E19 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | MR vessel wall imaging (VWI) is increasingly performed in clinical settings to support treatment decision-making regarding intracranial aneurysms. Aneurysm wall enhancement after contrast agent injection is expected to be related to aneurysm instability and rupture status. However, the authors hypothesize that slow-flow artifacts mimic aneurysm wall enhancement. Therefore, in this phantom study they assess the effect of slow flow on wall-like enhancement by using different MR VWI techniques.
The authors developed an MR-compatible aneurysm phantom model, which was connected to a pump to enable pulsatile inflow conditions. For VWI, 3D turbo spin echo sequences-both with and without motion-sensitized driven equilibrium (MSDE) and delay alternating with nutation for tailored excitation (DANTE) preparation pulses-were performed using a 3-T MR scanner. VWI was acquired both before and after Gd contrast agent administration by using two different pulsatile inflow conditions (2.5 ml/sec peak flow at 77 and 48 beats per minute). The intraluminal signal intensity along the aneurysm wall was analyzed to assess the performance of slow-flow suppression.
The authors observed wall-like enhancement after contrast agent injection, especially in low pump rate settings. Preparation pulses, in particular the DANTE technique, improved the performance of slow-flow suppression.
Near-wall slow flow mimics wall enhancement in VWI protocols. Therefore, VWI should be carefully interpreted. Preparation pulses improve slow-flow suppression, and therefore the authors encourage further development and clinical implementation of these techniques. |
---|---|