A mechanistic investigation of the EDWARDS INTUITY Elite valve’s hemodynamic performance
Objective Rapid deployment surgical aortic valve replacement has emerged as an alternative to the contemporary sutured valve technique. A difference in transvalvular pressure has been observed clinically between RD-SAVR and contemporary SAVR. A mechanistic inquiry into the impact of the rapid deploy...
Gespeichert in:
Veröffentlicht in: | General thoracic and cardiovascular surgery 2020-01, Vol.68 (1), p.9-17 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
Rapid deployment surgical aortic valve replacement has emerged as an alternative to the contemporary sutured valve technique. A difference in transvalvular pressure has been observed clinically between RD-SAVR and contemporary SAVR. A mechanistic inquiry into the impact of the rapid deployment valve inflow frame design on the left ventricular outflow tract and valve hemodynamics is needed.
Methods
A 23 mm EDWARDS INTUITY Elite rapid deployment valve and a control contemporary, sutured valve, a 23 mm Magna Ease valve, were implanted in an explanted human heart by an experienced cardiac surgeon. Per convention, the rapid deployment valve was implanted with three non-pledgeted, simple guiding sutures, while fifteen pledgeted, mattress sutures were used to implant the contemporary surgical valve. In vitro flow models were created from micro-computed tomography scans of the implanted valves and surrounding cardiac anatomy. Particle image velocimetry and hydrodynamic characterization experiments were conducted in the vicinity of the valves in a validated pulsatile flow loop system.
Results
The rapid deployment and control valves were found to have mean transvalvular pressure gradients of 7.92 ± 0.37 and 10.13 ± 0.48 mmHg, respectively. The inflow frame of the rapid deployment valve formed a larger, more circular, left ventricular outflow tract compared to the control valve. Furthermore, it was found that the presence of the control valve’s sub-annular pledgets compromised its velocity distribution and consequently its pressure gradient.
Conclusions
The rapid deployment valve’s intra-annular inflow frame provides for a larger, left ventricular outflow tract, thus reducing the transvalvular pressure gradient and improving overall hemodynamic performance. |
---|---|
ISSN: | 1863-6705 1863-6713 |