Comparison of magnetization transfer contrast of conventional and simultaneous multislice turbo spin echo acquisitions focusing on excitation time interval
Purpose Image contrast differs between conventional multislice turbo spin echo (conventional TSE) and multiband turbo spin echo (SMS-TSE). Difference in time interval between excitations for adjacent slices (SETI) might cause this difference. This study aimed to evaluate the influence of SETI on MT...
Gespeichert in:
Veröffentlicht in: | Japanese journal of radiology 2019-08, Vol.37 (8), p.579-589 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Image contrast differs between conventional multislice turbo spin echo (conventional TSE) and multiband turbo spin echo (SMS-TSE). Difference in time interval between excitations for adjacent slices (SETI) might cause this difference. This study aimed to evaluate the influence of SETI on MT effect for conventional TSE and compare conventional TSE with SMS-TSE in this respect.
Materials and methods
Three different agar concentration phantoms were scanned with conventional TSE by adjusting SETI and TR. Signal change for different SETI was evaluated using Pearson’s correlation analysis. SMS-TSE was acquired by changing TR similarly. Three human volunteers were scanned with similar settings to evaluate reproducibility of the phantom results in human brain.
Results
In conventional TSE, shorter SETI induced larger signal reduction. Longer TR and higher agar concentration emphasized this characteristic. Significant linear correlation (
P
|
---|---|
ISSN: | 1867-1071 1867-108X |
DOI: | 10.1007/s11604-019-00848-w |