CdTe quantum dots coated with a molecularly imprinted polymer for fluorometric determination of norfloxacin in seawater
A fluorescent nanoprobe consisting of CdTe quantum dots (QDs) and coated with molecularly imprinted layers was prepared and successfully applied to the determination of norfloxacin (NOR) in seawater and wastewater samples. The 3-mercaptopropionic acid capped QDs were prepared and then covered with a...
Gespeichert in:
Veröffentlicht in: | Mikrochimica acta (1966) 2019-06, Vol.186 (6), p.362-362, Article 362 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A fluorescent nanoprobe consisting of CdTe quantum dots (QDs) and coated with molecularly imprinted layers was prepared and successfully applied to the determination of norfloxacin (NOR) in seawater and wastewater samples. The 3-mercaptopropionic acid capped QDs were prepared and then covered with a protective silica shell. A molecularly imprinted layer was finally synthesized around the silanized QDs using 3-aminopropyltriethoxysilane as functional monomer and norfloxacin as the template. Compared with the non-imprinted polymer, the fluorescence of the nanoprobe with imprinted polymer (measured at excitation/emission wavelengths of 300/596 nm) is strongly reduced in the presence of NOR, and the imprinting factor is 8.8. Under the optimal experimental conditions, the detection limit of the nanoprobe is 0.18 μM, and response is linear between 0.5 - 28 μM of NOR. The relative standard deviation of the detection of NOR is |
---|---|
ISSN: | 0026-3672 1436-5073 |
DOI: | 10.1007/s00604-019-3440-7 |