Urgent Delivery - Validation and Operational Implementation of Urgent Blood Delivery by Modern High Speed Hospital Pneumatic Tube System to Support Bleeding Emergencies Within a Hospital Massive Transfusion Protocol

Timely blood delivery to patients with critical bleeding poses logistic challenges. A modern, high speed hospital pneumatic tube system (PTS) is one solution, but blood units may be subjected to high-speed torque and acceleration/deceleration forces. To validate a new PTS system for potential use at...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Laboratory medicine 2019-11, Vol.50 (4), p.e59-e69
Hauptverfasser: Chay, Jason Wai Mun, Kim, Ji Hyun, B M Taha, Nur Shahirah, Chiew, Siew Hong, Chin, Kwong Weng, Lim, Shu Hong, Ho, Liam Pock
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Timely blood delivery to patients with critical bleeding poses logistic challenges. A modern, high speed hospital pneumatic tube system (PTS) is one solution, but blood units may be subjected to high-speed torque and acceleration/deceleration forces. To validate a new PTS system for potential use at our 1,400-bed hospital in Singapore. Our validation included red blood cells, platelets, thawed plasma, and cryoprecipitate units transported from the blood bank for a distance of 820 meters (PTS track), at a velocity of 3-6 meters per second. Transit time, temperature, bag integrity, and blood quality were assessed visually and through analytical testing on pre- and post-PTS specimens. Blood units arrived physically intact in less than 8 minutes. The temperature for each was within the acceptable range. Comparative testing of pre-PTS and post-PTS specimens showed no significant difference in physical quality and analyzed parameters (P> .05). High speed PTS transportation of blood components has satisfactory fidelity and speed, without significant impact on quality. As a result, we incorporated PTS blood delivery into the hospital massive-transfusion protocol and successfully operationalized that new system.
ISSN:0007-5027
1943-7730
DOI:10.1093/labmed/lmz012