Insights from atrial surface activation throughout atrial tachycardia cycle length: A new mapping tool
A novel “LUMIPOINT” software in the Rhythmia system (Boston Scientific) displays a histogram of activated area over the entire atrial tachycardia (AT) cycle length (CL) with a normalized score. The purpose of this study was to examine whether the pattern of this global activation histogram (GAH) ide...
Gespeichert in:
Veröffentlicht in: | Heart rhythm 2019-11, Vol.16 (11), p.1652-1660 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A novel “LUMIPOINT” software in the Rhythmia system (Boston Scientific) displays a histogram of activated area over the entire atrial tachycardia (AT) cycle length (CL) with a normalized score.
The purpose of this study was to examine whether the pattern of this global activation histogram (GAH) identified reentrant vs focal AT and whether a decrease in atrial activation area, shown as valleys in the GAH, identifies isthmuses.
One hundred eight activation maps of ATs (17 focal, 57 macroreentrant, 21 localized, 13 multiple loop) in 67 patients were reviewed retrospectively with the LUMIPOINT software. The ACTIVATION SEARCH feature highlighted the activated area in a given time period irrespective of the activation map. A 30-ms unit time interval was set, and the GAH patterns and electrophysiological properties of highlighted areas were examined.
Focal ATs systematically displayed a plateau with GAH-Score |
---|---|
ISSN: | 1547-5271 1556-3871 |
DOI: | 10.1016/j.hrthm.2019.04.029 |