Correlation between incremental remifentanil doses and the Nociception Level (NOL) index response after intraoperative noxious stimuli
Background The Pain Monitoring Device (PMD) monitor (Medasense Biometrics Ltd., Ramat Gan, Israel) uses the Nociception Level (NOL) index, a multiple parameter-derived index that has recently shown a good sensitivity and specificity to detect noxious stimuli. The aim of this study was to assess the...
Gespeichert in:
Veröffentlicht in: | Canadian journal of anesthesia 2019-09, Vol.66 (9), p.1049-1061 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The Pain Monitoring Device (PMD) monitor (Medasense Biometrics Ltd., Ramat Gan, Israel) uses the Nociception Level (NOL) index, a multiple parameter-derived index that has recently shown a good sensitivity and specificity to detect noxious stimuli. The aim of this study was to assess the latest version of the device (PMD200
TM
) on variations of the NOL response after standardized tetanic stimuli to study the correlation between remifentanil doses and NOL.
Methods
Data from 26 patients undergoing midline laparotomy and receiving a desflurane-remifentanil-based anesthetic coupled with low thoracic epidural analgesia were analyzed. A standardized tetanic stimulus was applied to the forearm of the patients at different remifentanil infusion rates. The primary aim was to evaluate the correlation between post-tetanic stimulation NOL values from the PMD200 and remifentanil doses. The NOL index variations after experimental and clinical stimuli were also compared with heart rate (HR), mean arterial pressure (MAP), and Bispectral Index™ (BIS).
Results
A correlation between post-tetanic stimulation NOL values and remifentanil doses was found (r = −0.56; 95% confidence interval [CI], −0.70 to −0.44;
P
< 0.001). The NOL discriminated noxious from non-noxious states with the maximal Youden’s index value of the NOL receiver operating characteristic (ROC) curve showing a specificity of 88% (95% CI, 69.0 to 100) and sensitivity of 79.1% (95% CI, 56.2 to 95.5). The area under the NOL ROC curve (AUC, 0.9; 95% CI, 0.84 to 0.95) was significantly different from the other variables (
P
< 0.001
vs
HR;
P
< 0.001
vs
MAP;
P
< 0.001
vs
BIS).
Conclusions
The NOL value after noxious stimulus decreased with incremental remifentanil doses, showing a significant inverse correlation between the NOL index and opioid doses. The sensitivity and specificity of NOL to discriminate between noxious and non-noxious stimuli suggests its interesting potential as a monitor of nociception intensity during anesthesia.
Trial registration
www.clinicaltrials.gov
(NCT 02884778); 27 July, 2016. |
---|---|
ISSN: | 0832-610X 1496-8975 |
DOI: | 10.1007/s12630-019-01372-1 |