Protocol optimization for cardiac and liver iron content assessment using MRI: What sequence should I use?
To determine the optimal MRI protocol and sequences for liver and cardiac iron estimation in children. We evaluated patients ≤18 years with cardiac and liver MRIs for iron content estimation. Liver T2 was determined by a third-party company. Cardiac and Liver T2* values were measured by an observer....
Gespeichert in:
Veröffentlicht in: | Clinical imaging 2019-07, Vol.56, p.52-57 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To determine the optimal MRI protocol and sequences for liver and cardiac iron estimation in children.
We evaluated patients ≤18 years with cardiac and liver MRIs for iron content estimation. Liver T2 was determined by a third-party company. Cardiac and Liver T2* values were measured by an observer. Liver T2* values were calculated using the available liver parenchyma in the cardiac MRI. Linear correlations and Bland-Altman plots were run between liver T2 and T2*, cardiac T2* values; and liver T2* on dedicated cardiac and liver MRIs.
139 patients were included. Mean liver T2 and T2* values were 8.6 ± 5.4 ms and 4.5 ± 4.1 ms, respectively. A strong correlation between liver T2 and T2* values was observed (r = 0.96, p |
---|---|
ISSN: | 0899-7071 1873-4499 |