Sample matrix and high-sensitivity cardiac troponin I assays

Background Manufacturers of high-sensitivity cardiac troponin (hs-cTn) assays have restricted use of what sample types or matrices are acceptable to use for measurement. Our goal was to evaluate the comparability of the Siemens ADVIA Centaur hs-cTnI assay across different matrices and under differen...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical chemistry and laboratory medicine 2019-04, Vol.57 (5), p.745-751
Hauptverfasser: Kavsak, Peter A., Roy, Chantele, Malinowski, Paul, Clark, Lorna, Lamers, Shana, Bamford, Karen, Hill, Stephen, Worster, Andrew, Jaffe, Allan S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background Manufacturers of high-sensitivity cardiac troponin (hs-cTn) assays have restricted use of what sample types or matrices are acceptable to use for measurement. Our goal was to evaluate the comparability of the Siemens ADVIA Centaur hs-cTnI assay across different matrices and under different storage conditions. Methods Three different QC-plasma matrices were evaluated for imprecision