Histopathologic differences in the endovenous laser ablation between jacketed and radial fibers, in an ex vivo dominant extrafascial tributary of the great saphenous vein in an in vitro model, using histology and immunohistochemistry
The study aimed to investigate the biologic effects of the 1470-nm endovenous laser (EVL), with a jacketed fiber and a radial fiber, during EVL ablation of an ex vivo dominant extrafascial tributary of the great saphenous vein in our in vitro model by histology and immunohistochemistry. Ten segments...
Gespeichert in:
Veröffentlicht in: | Journal of vascular surgery. Venous and lymphatic disorders (New York, NY) NY), 2019-03, Vol.7 (2), p.234-245 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The study aimed to investigate the biologic effects of the 1470-nm endovenous laser (EVL), with a jacketed fiber and a radial fiber, during EVL ablation of an ex vivo dominant extrafascial tributary of the great saphenous vein in our in vitro model by histology and immunohistochemistry.
Ten segments of the dominant extrafascial tributary of the great saphenous vein were harvested by a consultant vascular surgeon from patients during routine varicose vein surgery. Six segments were treated using an ex vivo model of our design by a 1470-nm EVL with a jacketed fiber. The other four segments were also treated by a 1470-nm EVL but with a radial-firing fiber. Each segment was split into five sections and treated at five different linear endovenous energy densities (LEEDs) at 10 W: 0, 20, 40, 60, and 80 J/cm. The veins were incubated and subsections collected at 6 and 24 hours after treatment. Subsections were immersed in buffered formalin and taken for histologic and immunohistochemical analysis. Histopathologic analysis was then performed.
Treatment with the radial fiber led to a pattern of damage that was more homogeneous than with the jacketed fiber, with no carbonization of tissue present. Significant transmural damage and necrosis were observed at LEEDs of 60 and 80 J/cm in both treatment groups. At the same LEEDs, p53 and caspase 3 analysis showed that transmural cell wall vein death (necrosis or apoptosis) occurred by 6 hours after treatment with both fibers.
There was a significant difference in the effects of treatment with a jacketed fiber and a radial fiber in EVL ablation in vitro. Although both fibers caused transmural vein wall cell death at similar LEEDs, the pattern of damage with the radial fiber was more homogeneous. There was no overtreatment of tissue in terms of carbonization after treatment with the radial fiber. Treatment with the jacketed fiber showed carbonization of tissue at the same LEEDs.
Endovenous laser ablation has become one of the first-line choices for endovenous thermal ablation of incompetent truncal veins in modern varicose vein surgery. The 1470-nm laser targeted at water as the chromophore has become the “industry standard” for endovenous laser ablation. Currently, the two most commonly used devices are the jacketed fiber and the radial fiber. The jacketed fiber transmits the laser energy as a divergent cone in a forward-firing manner from the tip of the laser fiber, whereas the radial fiber directs the laser beam circumfe |
---|---|