Evaluation of a Ferromagnetic Marker Technology for Intraoperative Localization of Nonpalpable Breast Lesions
The purpose of this study was to evaluate the magnetic occult lesion localization instrument (MOLLI) system that involves implantation of a small, ferromagnetic marker to guide surgical excision of nonpalpable breast lesions. Characterization of the system was undertaken as part of what is, to our k...
Gespeichert in:
Veröffentlicht in: | American journal of roentgenology (1976) 2019-04, Vol.212 (4), p.727-733 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 733 |
---|---|
container_issue | 4 |
container_start_page | 727 |
container_title | American journal of roentgenology (1976) |
container_volume | 212 |
creator | Nicolae, Alexandru Dillon, John Semple, Mark Hong, Nicole Look Ravi, Ananth |
description | The purpose of this study was to evaluate the magnetic occult lesion localization instrument (MOLLI) system that involves implantation of a small, ferromagnetic marker to guide surgical excision of nonpalpable breast lesions. Characterization of the system was undertaken as part of what is, to our knowledge, the first study to assess the MOLLI system.
The MOLLI system consists of a handheld probe that can detect the position and distance of an implanted magnetic marker. The system presents the surgeon with an accurate assessment of lesion location and depth measurement for precise 3D localization. The marker is implanted under ultrasound or mammographic guidance at any time before the surgical procedure and requires no special precautions. Experimental analysis focused on characterization of the following aspects of the MOLLI system: visualization of the marker under imaging, 3D detection of the magnetic marker, spatial resolution of the probe to detect markers placed in close proximity, and the effect of signal interference on system performance.
The MOLLI system can reliably detect mean (± SD) marker depths up to 53 ± 8.56 mm from the probe. Bracketing large lesions or localizing multiple lesions can be accomplished by placing markers as close as 10 mm apart, at depths of up to 42 mm. The biologically inert MOLLI marker is readily visible under ultrasound and mammographic guidance, and it is differentiable from radiologic clips. The effect of surgical instruments on MOLLI functioning is minimal and does not impact system accuracy or reliability.
The MOLLI system offers an accurate and efficient alternative lesion localization method for nonpalpable breast lesions. |
doi_str_mv | 10.2214/AJR.18.20195 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2179453831</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2179453831</sourcerecordid><originalsourceid>FETCH-LOGICAL-c291t-42992cc3f47c4f73445f404b8917394e04b27a8c1d599dc599b879f1d06bc1033</originalsourceid><addsrcrecordid>eNo9kDFPwzAQRi0EoqWwMSOPDKT4YieOx1K1UFRAQkViixzHLoEkDnZSqfx6ApQudyd9777hIXQOZByGwK4n989jSMYhAREdoCFELA4oMDhEQ0JjCBJCXwfoxPt3QghPBD9GA0piTimjQ1TNNrLsZFvYGluDJZ5r52wl17VuC4UfpPvQDq-0eqttaddbbKzDi7p10jba9X8bjZdWybL42pc82rqRZSOzUuMbp6Vv8VL7PvSn6MjI0uuz3R6hl_lsNb0Llk-3i-lkGahQQBuwUIhQKWoYV8xwylhkGGFZIoBTwXR_hlwmCvJIiFz1I0u4MJCTOFNAKB2hy7_extnPTvs2rQqvdFnKWtvOpyFwwSKaUOjRqz9UOeu90yZtXFFJt02BpD-C015wCkn6K7jHL3bNXVbpfA__G6Xfvld2hQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2179453831</pqid></control><display><type>article</type><title>Evaluation of a Ferromagnetic Marker Technology for Intraoperative Localization of Nonpalpable Breast Lesions</title><source>American Roentgen Ray Society</source><source>MEDLINE</source><source>Alma/SFX Local Collection</source><creator>Nicolae, Alexandru ; Dillon, John ; Semple, Mark ; Hong, Nicole Look ; Ravi, Ananth</creator><creatorcontrib>Nicolae, Alexandru ; Dillon, John ; Semple, Mark ; Hong, Nicole Look ; Ravi, Ananth</creatorcontrib><description>The purpose of this study was to evaluate the magnetic occult lesion localization instrument (MOLLI) system that involves implantation of a small, ferromagnetic marker to guide surgical excision of nonpalpable breast lesions. Characterization of the system was undertaken as part of what is, to our knowledge, the first study to assess the MOLLI system.
The MOLLI system consists of a handheld probe that can detect the position and distance of an implanted magnetic marker. The system presents the surgeon with an accurate assessment of lesion location and depth measurement for precise 3D localization. The marker is implanted under ultrasound or mammographic guidance at any time before the surgical procedure and requires no special precautions. Experimental analysis focused on characterization of the following aspects of the MOLLI system: visualization of the marker under imaging, 3D detection of the magnetic marker, spatial resolution of the probe to detect markers placed in close proximity, and the effect of signal interference on system performance.
The MOLLI system can reliably detect mean (± SD) marker depths up to 53 ± 8.56 mm from the probe. Bracketing large lesions or localizing multiple lesions can be accomplished by placing markers as close as 10 mm apart, at depths of up to 42 mm. The biologically inert MOLLI marker is readily visible under ultrasound and mammographic guidance, and it is differentiable from radiologic clips. The effect of surgical instruments on MOLLI functioning is minimal and does not impact system accuracy or reliability.
The MOLLI system offers an accurate and efficient alternative lesion localization method for nonpalpable breast lesions.</description><identifier>ISSN: 0361-803X</identifier><identifier>EISSN: 1546-3141</identifier><identifier>DOI: 10.2214/AJR.18.20195</identifier><identifier>PMID: 30673343</identifier><language>eng</language><publisher>United States</publisher><subject>Breast Neoplasms - diagnostic imaging ; Breast Neoplasms - surgery ; Coated Materials, Biocompatible ; Equipment Design ; Female ; Fiducial Markers ; Humans ; Imaging, Three-Dimensional ; Magnetics - instrumentation ; Mammography ; Needles ; Radiography, Interventional ; Surgery, Computer-Assisted - instrumentation ; Ultrasonography, Interventional ; Ultrasonography, Mammary</subject><ispartof>American journal of roentgenology (1976), 2019-04, Vol.212 (4), p.727-733</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c291t-42992cc3f47c4f73445f404b8917394e04b27a8c1d599dc599b879f1d06bc1033</citedby><cites>FETCH-LOGICAL-c291t-42992cc3f47c4f73445f404b8917394e04b27a8c1d599dc599b879f1d06bc1033</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,4110,27915,27916</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/30673343$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Nicolae, Alexandru</creatorcontrib><creatorcontrib>Dillon, John</creatorcontrib><creatorcontrib>Semple, Mark</creatorcontrib><creatorcontrib>Hong, Nicole Look</creatorcontrib><creatorcontrib>Ravi, Ananth</creatorcontrib><title>Evaluation of a Ferromagnetic Marker Technology for Intraoperative Localization of Nonpalpable Breast Lesions</title><title>American journal of roentgenology (1976)</title><addtitle>AJR Am J Roentgenol</addtitle><description>The purpose of this study was to evaluate the magnetic occult lesion localization instrument (MOLLI) system that involves implantation of a small, ferromagnetic marker to guide surgical excision of nonpalpable breast lesions. Characterization of the system was undertaken as part of what is, to our knowledge, the first study to assess the MOLLI system.
The MOLLI system consists of a handheld probe that can detect the position and distance of an implanted magnetic marker. The system presents the surgeon with an accurate assessment of lesion location and depth measurement for precise 3D localization. The marker is implanted under ultrasound or mammographic guidance at any time before the surgical procedure and requires no special precautions. Experimental analysis focused on characterization of the following aspects of the MOLLI system: visualization of the marker under imaging, 3D detection of the magnetic marker, spatial resolution of the probe to detect markers placed in close proximity, and the effect of signal interference on system performance.
The MOLLI system can reliably detect mean (± SD) marker depths up to 53 ± 8.56 mm from the probe. Bracketing large lesions or localizing multiple lesions can be accomplished by placing markers as close as 10 mm apart, at depths of up to 42 mm. The biologically inert MOLLI marker is readily visible under ultrasound and mammographic guidance, and it is differentiable from radiologic clips. The effect of surgical instruments on MOLLI functioning is minimal and does not impact system accuracy or reliability.
The MOLLI system offers an accurate and efficient alternative lesion localization method for nonpalpable breast lesions.</description><subject>Breast Neoplasms - diagnostic imaging</subject><subject>Breast Neoplasms - surgery</subject><subject>Coated Materials, Biocompatible</subject><subject>Equipment Design</subject><subject>Female</subject><subject>Fiducial Markers</subject><subject>Humans</subject><subject>Imaging, Three-Dimensional</subject><subject>Magnetics - instrumentation</subject><subject>Mammography</subject><subject>Needles</subject><subject>Radiography, Interventional</subject><subject>Surgery, Computer-Assisted - instrumentation</subject><subject>Ultrasonography, Interventional</subject><subject>Ultrasonography, Mammary</subject><issn>0361-803X</issn><issn>1546-3141</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9kDFPwzAQRi0EoqWwMSOPDKT4YieOx1K1UFRAQkViixzHLoEkDnZSqfx6ApQudyd9777hIXQOZByGwK4n989jSMYhAREdoCFELA4oMDhEQ0JjCBJCXwfoxPt3QghPBD9GA0piTimjQ1TNNrLsZFvYGluDJZ5r52wl17VuC4UfpPvQDq-0eqttaddbbKzDi7p10jba9X8bjZdWybL42pc82rqRZSOzUuMbp6Vv8VL7PvSn6MjI0uuz3R6hl_lsNb0Llk-3i-lkGahQQBuwUIhQKWoYV8xwylhkGGFZIoBTwXR_hlwmCvJIiFz1I0u4MJCTOFNAKB2hy7_extnPTvs2rQqvdFnKWtvOpyFwwSKaUOjRqz9UOeu90yZtXFFJt02BpD-C015wCkn6K7jHL3bNXVbpfA__G6Xfvld2hQ</recordid><startdate>201904</startdate><enddate>201904</enddate><creator>Nicolae, Alexandru</creator><creator>Dillon, John</creator><creator>Semple, Mark</creator><creator>Hong, Nicole Look</creator><creator>Ravi, Ananth</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>201904</creationdate><title>Evaluation of a Ferromagnetic Marker Technology for Intraoperative Localization of Nonpalpable Breast Lesions</title><author>Nicolae, Alexandru ; Dillon, John ; Semple, Mark ; Hong, Nicole Look ; Ravi, Ananth</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c291t-42992cc3f47c4f73445f404b8917394e04b27a8c1d599dc599b879f1d06bc1033</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><topic>Breast Neoplasms - diagnostic imaging</topic><topic>Breast Neoplasms - surgery</topic><topic>Coated Materials, Biocompatible</topic><topic>Equipment Design</topic><topic>Female</topic><topic>Fiducial Markers</topic><topic>Humans</topic><topic>Imaging, Three-Dimensional</topic><topic>Magnetics - instrumentation</topic><topic>Mammography</topic><topic>Needles</topic><topic>Radiography, Interventional</topic><topic>Surgery, Computer-Assisted - instrumentation</topic><topic>Ultrasonography, Interventional</topic><topic>Ultrasonography, Mammary</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Nicolae, Alexandru</creatorcontrib><creatorcontrib>Dillon, John</creatorcontrib><creatorcontrib>Semple, Mark</creatorcontrib><creatorcontrib>Hong, Nicole Look</creatorcontrib><creatorcontrib>Ravi, Ananth</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>American journal of roentgenology (1976)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Nicolae, Alexandru</au><au>Dillon, John</au><au>Semple, Mark</au><au>Hong, Nicole Look</au><au>Ravi, Ananth</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Evaluation of a Ferromagnetic Marker Technology for Intraoperative Localization of Nonpalpable Breast Lesions</atitle><jtitle>American journal of roentgenology (1976)</jtitle><addtitle>AJR Am J Roentgenol</addtitle><date>2019-04</date><risdate>2019</risdate><volume>212</volume><issue>4</issue><spage>727</spage><epage>733</epage><pages>727-733</pages><issn>0361-803X</issn><eissn>1546-3141</eissn><abstract>The purpose of this study was to evaluate the magnetic occult lesion localization instrument (MOLLI) system that involves implantation of a small, ferromagnetic marker to guide surgical excision of nonpalpable breast lesions. Characterization of the system was undertaken as part of what is, to our knowledge, the first study to assess the MOLLI system.
The MOLLI system consists of a handheld probe that can detect the position and distance of an implanted magnetic marker. The system presents the surgeon with an accurate assessment of lesion location and depth measurement for precise 3D localization. The marker is implanted under ultrasound or mammographic guidance at any time before the surgical procedure and requires no special precautions. Experimental analysis focused on characterization of the following aspects of the MOLLI system: visualization of the marker under imaging, 3D detection of the magnetic marker, spatial resolution of the probe to detect markers placed in close proximity, and the effect of signal interference on system performance.
The MOLLI system can reliably detect mean (± SD) marker depths up to 53 ± 8.56 mm from the probe. Bracketing large lesions or localizing multiple lesions can be accomplished by placing markers as close as 10 mm apart, at depths of up to 42 mm. The biologically inert MOLLI marker is readily visible under ultrasound and mammographic guidance, and it is differentiable from radiologic clips. The effect of surgical instruments on MOLLI functioning is minimal and does not impact system accuracy or reliability.
The MOLLI system offers an accurate and efficient alternative lesion localization method for nonpalpable breast lesions.</abstract><cop>United States</cop><pmid>30673343</pmid><doi>10.2214/AJR.18.20195</doi><tpages>7</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0361-803X |
ispartof | American journal of roentgenology (1976), 2019-04, Vol.212 (4), p.727-733 |
issn | 0361-803X 1546-3141 |
language | eng |
recordid | cdi_proquest_miscellaneous_2179453831 |
source | American Roentgen Ray Society; MEDLINE; Alma/SFX Local Collection |
subjects | Breast Neoplasms - diagnostic imaging Breast Neoplasms - surgery Coated Materials, Biocompatible Equipment Design Female Fiducial Markers Humans Imaging, Three-Dimensional Magnetics - instrumentation Mammography Needles Radiography, Interventional Surgery, Computer-Assisted - instrumentation Ultrasonography, Interventional Ultrasonography, Mammary |
title | Evaluation of a Ferromagnetic Marker Technology for Intraoperative Localization of Nonpalpable Breast Lesions |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-14T23%3A14%3A56IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Evaluation%20of%20a%20Ferromagnetic%20Marker%20Technology%20for%20Intraoperative%20Localization%20of%20Nonpalpable%20Breast%20Lesions&rft.jtitle=American%20journal%20of%20roentgenology%20(1976)&rft.au=Nicolae,%20Alexandru&rft.date=2019-04&rft.volume=212&rft.issue=4&rft.spage=727&rft.epage=733&rft.pages=727-733&rft.issn=0361-803X&rft.eissn=1546-3141&rft_id=info:doi/10.2214/AJR.18.20195&rft_dat=%3Cproquest_cross%3E2179453831%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2179453831&rft_id=info:pmid/30673343&rfr_iscdi=true |