The Application of a New Model-Based Iterative Reconstruction in Low-Dose Upper Abdominal CT

To compare upper abdominal computed tomography (CT) image quality of new model-based iterative reconstruction (MBIR) with low-contrast resolution preference (MBIRNR40), conventional MBIR (MBIRc), and adaptive statistical iterative reconstruction (ASIR) at low dose with ASIR at routine-dose. Study in...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Academic radiology 2019-10, Vol.26 (10), p.e275-e283
Hauptverfasser: Jia, Yongjun, Zhai, Bingying, He, Taiping, Yu, Yong, Yu, Nan, Duan, Haifeng, Yang, Chuangbo, Zhang, Xirong
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To compare upper abdominal computed tomography (CT) image quality of new model-based iterative reconstruction (MBIR) with low-contrast resolution preference (MBIRNR40), conventional MBIR (MBIRc), and adaptive statistical iterative reconstruction (ASIR) at low dose with ASIR at routine-dose. Study included phantom and 60 patients who had initial and follow-up CT scans. For patients, the delay phase was acquired at routine-dose (noise index = 10 HU) for the initial scan and low dose (noise index = 20 HU) for the follow-up. The low-dose CT was reconstructed with 40% and 60% ASIR, MBIRc, and MBIRNR40, while routine-dose CT was reconstructed with 40% ASIR. CT value and noise measurements of the subcutaneous fat, back muscle, liver, and spleen parenchyma were compared using one-way ANOVA. Two radiologists used semiquantitative 7-scale (−3 to +3) to rate image quality and artifacts. The phantom study revealed superior low-contrast resolution with MBIRNR40. For patient scans, the CT dose index for the low-dose CT was 3.00 ± 1.32 mGy, 75% lower than the 11.90 ± 4.75 mGy for the routine-dose CT. Image noise for the low-dose MBIRNR40 images was significantly lower than the low-dose MBIRc and ASIR images, and routine-dose ASIR images (p < 0.05). Subjective ratings showed higher image quality for low-dose MBIRNR40, with lower noise, better low-contrast resolution for abdominal structures, and finer lesion contours than those of low-dose MBIRc and ASIR images, and routine-dose ASIR images (p < 0.05). MBIRNR40 with low-contrast resolution preference provides significantly lower noise and better image quality than MBIRc and ASIR in low-dose abdominal CT; significantly better objective and subjective image quality than the routine-dose ASIR with 75% dose reduction.
ISSN:1076-6332
1878-4046
DOI:10.1016/j.acra.2018.11.020