Characterization of steady-state wear in all-alumina THR to 20 Mc
The steady-state ceramic wear-rates was characterized to 20 million cycles test duration for the all-alumina total hip replacement (THR) as a follow-up to an earlier simulator study. Scanning electron microscopy (SEM) studies at 20 Mc showed that the original machine tracks were polished off and cra...
Gespeichert in:
Hauptverfasser: | , , , , |
---|---|
Format: | Tagungsbericht |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Schreiben Sie den ersten Kommentar!