Surface plasmon resonance biosensor based on integrated optical waveguide

We report a sensor based on spectral interrogation of surface plasmon resonance (SPR) in an integrated optical waveguide-coupled SPR sensing device. We present theoretical analysis of the integrated optical SPR sensor structure leading to a device design optimized for operation in aqueous environmen...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors and actuators. B, Chemical Chemical, 2001-06, Vol.76 (1), p.8-12
Hauptverfasser: Dostálek, J, Čtyroký, J, Homola, J, Brynda, E, Skalský, M, Nekvindová, P, Špirková, J, Škvor, J, Schröfel, J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We report a sensor based on spectral interrogation of surface plasmon resonance (SPR) in an integrated optical waveguide-coupled SPR sensing device. We present theoretical analysis of the integrated optical SPR sensor structure leading to a device design optimized for operation in aqueous environments. We demonstrate that the fabricated laboratory prototype of the sensor is capable of measuring bulk refractive index changes smaller than 1.2×10 −6. In conjunction with specific biomolecular recognition elements (monoclonal antibodies against human chorigonadotropin (hCG)) the sensor is used for the detection of hCG. The sensor is demonstrated to be capable of detecting 2 ng of hCG present in 1 ml of 1% bovine serum albumin solution.
ISSN:0925-4005
1873-3077
DOI:10.1016/S0925-4005(01)00559-7