Surface plasmon resonance biosensor based on integrated optical waveguide
We report a sensor based on spectral interrogation of surface plasmon resonance (SPR) in an integrated optical waveguide-coupled SPR sensing device. We present theoretical analysis of the integrated optical SPR sensor structure leading to a device design optimized for operation in aqueous environmen...
Gespeichert in:
Veröffentlicht in: | Sensors and actuators. B, Chemical Chemical, 2001-06, Vol.76 (1), p.8-12 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report a sensor based on spectral interrogation of surface plasmon resonance (SPR) in an integrated optical waveguide-coupled SPR sensing device. We present theoretical analysis of the integrated optical SPR sensor structure leading to a device design optimized for operation in aqueous environments. We demonstrate that the fabricated laboratory prototype of the sensor is capable of measuring bulk refractive index changes smaller than 1.2×10
−6. In conjunction with specific biomolecular recognition elements (monoclonal antibodies against human chorigonadotropin (hCG)) the sensor is used for the detection of hCG. The sensor is demonstrated to be capable of detecting 2
ng of hCG present in 1
ml of 1% bovine serum albumin solution. |
---|---|
ISSN: | 0925-4005 1873-3077 |
DOI: | 10.1016/S0925-4005(01)00559-7 |