Inhalable aerosol selector for the CIP 10 personal aerosol sampler

In this study, the CIP 10-I (Inhalable) aerosol sampler was used as a static sampler. The sampler exhibited a particle-size dependent sampling efficiency close to the conventional inhalable curve. Further, it was found that the sampling efficiency does not depend on the external wind speed.

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of aerosol science 1999-09, Vol.30 (Suppl. 1), p.S893-S894
Hauptverfasser: Görner, P., Wrobel, R., Roger, F., Fabriès, J.-F.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this study, the CIP 10-I (Inhalable) aerosol sampler was used as a static sampler. The sampler exhibited a particle-size dependent sampling efficiency close to the conventional inhalable curve. Further, it was found that the sampling efficiency does not depend on the external wind speed.
ISSN:0021-8502
1879-1964
DOI:10.1016/S0021-8502(99)80457-1