Wireless Real-Time Temperature Monitoring of Blood Packages: Silver Nanowire-Embedded Flexible Temperature Sensors

Real-time temperature monitoring of individual blood packages capable of wireless data transmission to ensure the safety of blood samples and minimize wastes has become a critical issue in recent years. In this work, we propose flexible temperature sensors using silver nanowires (NWs) and a flexible...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:ACS applied materials & interfaces 2018-12, Vol.10 (51), p.44678-44685
Hauptverfasser: Youn, Doo-Young, Jung, Uihyun, Naqi, Muhammad, Choi, Seon-Jin, Lee, Min-Goo, Lee, Sungho, Park, Hi-Joon, Kim, Il-Doo, Kim, Sunkook
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Real-time temperature monitoring of individual blood packages capable of wireless data transmission to ensure the safety of blood samples and minimize wastes has become a critical issue in recent years. In this work, we propose flexible temperature sensors using silver nanowires (NWs) and a flexible colorless polyimide (CPI) film integrated with a wireless data transmission circuit. The unique design of the temperature sensors was achieved by patterning Ag NWs using a three-dimensional printed mold and embedding the patterned Ag NWs in the CPI film (p-Ag NWs/CPI), which resulted in a flexible temperature sensor with electrical, mechanical, and temperature stability for applications in blood temperature monitoring. Indeed, a reliable resistance change of the p-Ag NWs/CPI was observed in the temperature range of −20 to 20 °C with a robust bending stability of up to 5000 cycles at 5 mm bending radius. Real-time and wireless temperature monitoring using the p-Ag NWs/CPI was demonstrated with the packages of rat blood. The result revealed that the stable and consistent temperature monitoring of individual blood packages could be achieved in a blood box, which was mainly attributed to the conformal attachment of the p-Ag NWs/CPI to different packages in a blood container.
ISSN:1944-8244
1944-8252
DOI:10.1021/acsami.8b11928