Virtual trial to evaluate the robustness of cementless femoral stems to patient and surgical variation

Primary stability is essential for the success of cementless femoral stems. In this study, patient specific finite element (FE) models were used to assess changes in primary stability due to variability in patient anatomy, bone properties and stem alignment for two commonly used cementless femoral s...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biomechanics 2019-01, Vol.82, p.346-356
Hauptverfasser: Al-Dirini, Rami M.A., Martelli, Saulo, O'Rourke, Dermot, Huff, Daniel, Zhang, Ju, Clement, John G., Besier, Thor, Taylor, Mark
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Primary stability is essential for the success of cementless femoral stems. In this study, patient specific finite element (FE) models were used to assess changes in primary stability due to variability in patient anatomy, bone properties and stem alignment for two commonly used cementless femoral stems, Corail® and Summit® (DePuy Synthes, Warsaw, USA). Computed-tomography images of the femur were obtained for 8 males and 8 females. An automated algorithm was used to determine the stem position and size which minimized the endo-cortical space, and then span the plausible surgical envelope of implant positions constrained by the endo-cortical boundary. A total of 1952 models were generated and ran, each with a unique alignment scenario. Peak hip contact and muscle forces for stair climbing were scaled to the donor’s body weight and applied to the model. The primary stability was assessed by comparing the implant micromotion and peri-prosthetic strains to thresholds (150 μm and 7000 µε, respectively) above which fibrous tissue differentiation and bone damage are expected to prevail. Despite the wide range of implant positions included, FE prediction were mostly below the thresholds (medians: Corail®: 20–74 µm and 1150–2884 µε, Summit®: 25–111 µm and 860–3010 µε), but sensitivity of micromotion and interfacial strains varied across femora, with the majority being sensitive (p 
ISSN:0021-9290