Ultrasonic measurement of sound velocity fluctuations in biological tissue due to ultrasonic heating and estimation of thermo-physical properties

Purpose Tissue characterization in terms of the differences in thermo-physical properties of biological tissues was investigated in this study. The objective was to measure the ratio of variation in sound velocity due to ultrasonic heating and to derive the relational expression between the ratio an...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of medical ultrasonics (2001) 2019-01, Vol.46 (1), p.35-43
Hauptverfasser: Tsujimoto, Yukako, Morimoto, Mai, Nitta, Naotaka, Akiyama, Iwaki
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose Tissue characterization in terms of the differences in thermo-physical properties of biological tissues was investigated in this study. The objective was to measure the ratio of variation in sound velocity due to ultrasonic heating and to derive the relational expression between the ratio and thermo-physical properties. Methods The ratio of sound velocity variation before and after the temperature rise of tissue samples exposed to ultrasound was measured by ultrasonic pulse echo method. The thermo-physical properties were estimated for a tissue-mimicking material and porcine muscle and fat tissues due to theoretical expression. The transducer for heating had a resonance frequency of 3.2 MHz, and the transducer for measurement of sound velocity variation had a resonance frequency of 5.2 MHz. Results In the phantom study, the measured values of the temperature rise agreed with the values calculated by a finite element method (FEM). The estimated values of the temperature rise from the sound velocity variations of muscle and fat tissues were 0.36 °C and 1.1 °C, respectively. Also, the estimated values of thermo-physical properties agreed with the reference values within an error of 10%. Conclusions The thermo-physical properties of the porcine tissues were measured by sound velocity variation due to ultrasonic heating within the safety regulations.
ISSN:1346-4523