Identification and detection sensitivity of Microcystis aeruginosa from mixed and field samples using MALDI-TOF MS
To verify the applicability of identifying Microcystis aeruginosa by matrix-assisted laser desorption-ionization-time-of-flight mass spectrometry (MALDI-TOF MS), mixed and field samples were employed to study the sensitivity and the analysis power, respectively. Series diluted samples and artificial...
Gespeichert in:
Veröffentlicht in: | Environmental monitoring and assessment 2018-12, Vol.190 (12), p.712-13, Article 712 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To verify the applicability of identifying
Microcystis aeruginosa
by matrix-assisted laser desorption-ionization-time-of-flight mass spectrometry (MALDI-TOF MS), mixed and field samples were employed to study the sensitivity and the analysis power, respectively. Series diluted samples and artificially mixed samples by the
M. aeruginosa
NIES-843 strain were designed to verify the sensitivity. The lowest detection limit was 1.955 × 10
6
cells in pure samples, while for mixed samples, the lowest detection limit and ratio of NIES-843 strain were 2.88 × 10
6
cells and 33.7%, respectively. The results provided a reference for the reasonable volume of the water sample in which the
M. aeruginosa
could be detected. Ribosomal protein biomarkers for identifying
M. aeruginosa
which were successfully detected from the field samples in Taihu Lake, indicated that the identification of
M. aeruginosa
by MALDI-TOF MS could be applied in field samples. Furthermore, different genetic types of
M. aeruginosa
strains were also detected at different locations in Taihu Lake, which revealed the diversity of
M. aeruginosa
and the detection power of MALDI-TOF MS at the strain level for the field samples. The sensitivity and detection power in the analysis of
M. aeruginosa
by the MALDI-TOF MS demonstrated the applicability of this method in routine environmental monitoring. |
---|---|
ISSN: | 0167-6369 1573-2959 |
DOI: | 10.1007/s10661-018-7093-5 |