Continuous differential monitoring of the spent dialysate glucose level: clinical evaluation
A new glucose monitoring system for the haemodialysis is presented. This new tool measures real-time glucose concentrations in the dialysate flow of the artificial kidney. The background signal coming from present clinical interferents is effectively rejected by using a differential measurement set-...
Gespeichert in:
Veröffentlicht in: | Sensors and actuators. B, Chemical Chemical, 1997-10, Vol.44 (1), p.304-308 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new glucose monitoring system for the haemodialysis is presented. This new tool measures real-time glucose concentrations in the dialysate flow of the artificial kidney. The background signal coming from present clinical interferents is effectively rejected by using a differential measurement set-up. The amperometric device is based on thick film RuO
2 sensors. They are connected to a novel portable system with an optimal patient safety protection. This is guaranteed by means of a galvanic-isolated power supply and an optically-isolated data-communication to a lap-top PC. The sensor interface can easily be controlled via a graphical user interface-screen. |
---|---|