Investigation on the resolution of a micro cone beam CT scanner scintillating detector using Monte Carlo methods
•Monte Carlo simulation of a scintillating pixel detector of a micro-CT scanner.•Modeling and quantification of physics processes of individual detector components.•Scatter radiation does not affect the spatial resolution of the detector.•The resolution is mainly limited due to the propagation of sc...
Gespeichert in:
Veröffentlicht in: | Physica medica 2018-09, Vol.53, p.17-24 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Monte Carlo simulation of a scintillating pixel detector of a micro-CT scanner.•Modeling and quantification of physics processes of individual detector components.•Scatter radiation does not affect the spatial resolution of the detector.•The resolution is mainly limited due to the propagation of scintillation light.
The impact of several physical quantities on the spatial resolution of an X-ray scintillating pixel detector for a micro cone beam CT (µCBCT) is investigated and discussed.
The XtremeCT from SCANCO Medical AG was simulated using the EGSnrc/EGS++ Monte Carlo (MC) framework and extensively benchmarked in a previous work. The resolution of the detector was determined by simulating a titanium knife-edge to obtain the edge spread function (ESF) and the modulation transfer function (MTF). Propagation of the scintillation light through the scintillator and its coupling into the fiber optics system was taken into account.
The contribution of particles scattered in the main scanner components to the detector signal is very low and does not affect the spatial resolution of the detector. The resolution obtained from the energy deposition in the scintillator without any blurring due to the propagation of the scintillation light into the fiber optics array was 31 µm. By assuming isotropic light propagation in the scintillator, the resolution degraded to 360 µm. A simple light propagation model taking into account the impact of the scintillator’s columnar microstructures was developed and compared with the MANTIS Monte Carlo simulation package. By reducing the width of the model’s light propagation kernel by a factor of 2 compared to the isotropic case, the detector resolution can be improved to 83 µm, which corresponds well to the measured resolution of 86 µm.
The resolution of the detector is limited mainly by the propagation of the scintillation light through the scintillator layer. It offers the greatest potential to improve the resolution of the µCBCT imaging system. |
---|---|