The performance of a multi guard ring (MGR) diode for clinical electron beams dosimetry
The dosimetric response of a multi guard ring structure (MGR) diode has been studied with clinical electron beam energies from 5 MeV to 15 MeV. The results showed that the MGR dose response is linear in the range of 5–320 cGy and presents reproducibility with variation coefficients less than 0.4%. T...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2018-11, Vol.141, p.112-117 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The dosimetric response of a multi guard ring structure (MGR) diode has been studied with clinical electron beam energies from 5 MeV to 15 MeV. The results showed that the MGR dose response is linear in the range of 5–320 cGy and presents reproducibility with variation coefficients less than 0.4%. The field output factors measured with the MGR agreed within 2% with those measured with an ionization chamber. This study evidences that this diode can be used for clinical electron beam dosimetry.
•Performance of a multi guard ring structure (MGR) for clinical electron beam dosimetry.•New type of diodes with enhanced tolerance to radiation is evaluated for electron dosimetry.•Clinical electron beam dosimetry with semiconductor detector. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2018.07.002 |