Intraocular Pressure Measurement Accuracy and Repeatability of a Modified Goldmann Prism: Multicenter Randomized Clinical Trial
To clinically evaluate a modified surface Goldmann applanation tomometer (GAT) prism for intraocular pressure (IOP) accuracy, repeatability, and safety. Prospective, open-label, randomized, controlled, multicenter reference device reliability and validity analysis. A GAT and a modified surface GAT p...
Gespeichert in:
Veröffentlicht in: | American journal of ophthalmology 2018-12, Vol.196, p.145-153 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To clinically evaluate a modified surface Goldmann applanation tomometer (GAT) prism for intraocular pressure (IOP) accuracy, repeatability, and safety.
Prospective, open-label, randomized, controlled, multicenter reference device reliability and validity analysis.
A GAT and a modified surface GAT prism measured IOP on 173 unique eyes. The study design and analysis complied with FDA 510(k) and ANSIZ80.10-2014 guidelines. All eyes were randomized to IOP measurement by 1 of 5 standard prisms or 5 modified prisms, each from a different manufacturing lot. Pressures were measured by 6 investigators, 2 times with each prism, for a total of 1384 IOP measurements. Analysis included Bland-Altman difference accuracy, intraoperator and interoperator IOP measurement, and manufactured lot repeatability.
Bland-Altman indicated no IOP measurements pairs outside the ±5 mm Hg guidelines. Operator and manufactured lot repeatability F tests and 1-way ANOVAs indicated no statistical difference between the standard and modified prisms (all P > .10). The difference in IOP measurements of the standard and modified prisms correlated well to Dresdner central corneal thickness (CCT) correction (P = .01).
A modified surface replacement prism is statistically equivalent to a flat-surfaced prism. The modified surface prism indicated statistically significant correction for CCT requiring further testing outside the ANSI standard limits (0.500 mm < CCT < 0.600 mm) to examine its full potential. |
---|---|
ISSN: | 0002-9394 1879-1891 1879-1891 |
DOI: | 10.1016/j.ajo.2018.08.051 |