The goal of this study was to examine a new weightless-like radiation protection garment regarding its radiation protection efficacy and to compare it to a conventional two-piece apron suit plus thyroid collar and standard ancillary shields. All measurements were carried out using a clinical angiogr...
Gespeichert in:
Veröffentlicht in: | European journal of radiology 2018-09, Vol.106, p.56-61 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The goal of this study was to examine a new weightless-like radiation protection garment regarding its radiation protection efficacy and to compare it to a conventional two-piece apron suit plus thyroid collar and standard ancillary shields.
All measurements were carried out using a clinical angiography system with a standardized fluoroscopy protocol for different C-arm angulations. An anthropomorphic torso phantom served as a scattering body. In addition, an ionization chamber was used to measure the radiation exposure on five different representative heights and at two different positions of an examiner during a typical fluoroscopic-guided intervention.
The new weightless-like radiation protection garment and the conventional protection concept showed a mean dose reduction of 98.1% (p |
---|---|
ISSN: | 0720-048X 1872-7727 |