Six-month evaluation of novel bioabsorbable scaffolds composed of poly-L-lactic acid and amorphous calcium phosphate nanoparticles in porcine coronary arteries
Objective Using coronary angiography and intravascular ultrasound methods to evaluate the performance of the novel fully bioabsorbable scaffold (NFBS) composed of poly-L-lactic acid/amorphous calcium phosphate (PLLA/ACP) at six-month follow-up by comparing with PLLA scaffolds Methods Twelve PLLA/ACP...
Gespeichert in:
Veröffentlicht in: | Journal of biomaterials applications 2018-08, Vol.33 (2), p.227-233 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
Using coronary angiography and intravascular ultrasound methods to evaluate the performance of the novel fully bioabsorbable scaffold (NFBS) composed of poly-L-lactic acid/amorphous calcium phosphate (PLLA/ACP) at six-month follow-up by comparing with PLLA scaffolds
Methods
Twelve PLLA/ACP scaffolds and 12 PLLA scaffolds were implanted into the coronary arteries of 12 miniature pigs. Quantitative coronary angiography (QCA) was used to measure the reference vessel diameter (RVD), mean lumen diameter (MLD) and late lumen loss (LLL). According to IVUS images, we calculated the strut malapposition rate (SMR) at post implantation, strut overlap rate (SOR), reference vessel area (RVA), mean stent area (MSA), mean lumen area (MLA) and luminal patency rate (LPR) at six-month follow-up. The radial strength of the scaffold was evaluated using a catheter tensile testing machine.
Results
QCA results indicated that, at six month, MLD of PLLA/ACP scaffolds was greater than those of PLLA scaffolds (2.47 ± 0.22 mm vs. 2.08 ± 0.25 mm, P |
---|---|
ISSN: | 0885-3282 1530-8022 |
DOI: | 10.1177/0885328218790332 |