Evaluation of free thyroxine determination based on one-step fluorometric immunoassay technique and the suboptimal concordance with two-step immunoassay

•AIA 900 system is suitable for routine diagnosis purpose.•AIA 900 was validated according to French protocol.•Low concordance between AIA900 and Access 2 immunoassays is reported.•Various clinical contexts may challenge the quantification of FT4. Free thyroxine (FT4) quantification is continuing to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of pharmaceutical and biomedical analysis 2018-10, Vol.160, p.187-194
Hauptverfasser: Doggui, Radhouene, Mahjoub, Rahma, Aissa, Wassef Lotfi, Tounsi, Imen, Allagui, Zeineb, Cherif, Zmorda, Haj Sassi, Sihem, Ennigrou, Sinda, Edriss, Dhiaeddine, El Ati, Jalila, Bibi, Amina
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:•AIA 900 system is suitable for routine diagnosis purpose.•AIA 900 was validated according to French protocol.•Low concordance between AIA900 and Access 2 immunoassays is reported.•Various clinical contexts may challenge the quantification of FT4. Free thyroxine (FT4) quantification is continuing to be a concern. The purpose of the following study was to evaluate the analytical performance of Tosoh AIA900 based on a one-step technique and its comparison to Access 2 (two-step technique) over different clinical contexts (euthyroid, thyroid disorders, uncontrolled diabetes, renal failure and pregnancy). The protocol established by the French society of Clinical Biology was used to evaluate: imprecision, limit of detection, trueness, linearity, interferences and method comparisons. Within-run variation of 3.1%, 5.7% and 4.4% were found for the low, medium and high controls, respectively. Between-run was 5.8% for low control, 5.7% for medium control and 7.1% for high control. Common interferences did not affect one-step immunoassay FT4 results. The linearity was checked up to 86 pmol/L. The limit of detection was 5.5 pmol/L. The concordance correlation coefficient (CCC) showed a low agreement (0.6) between both methods. Bland-Altman plot revealed that AIA 900 one-step immunoassay technique provides a significant higher values ((+2.8 ± 2.7 pmol/L;p 
ISSN:0731-7085
1873-264X
DOI:10.1016/j.jpba.2018.07.045