Technique of sonographic assessment of lower uterine segment in women with previous cesarean delivery: a prospective, pre/intraoperative comparative ultrasound study
Objective The purpose of this study was: (A) to establish the effects of different ultrasound measurement methods (linear versus curved array) and measuring conditions [impact of pressure by fetal head/pelvis on the lower uterine segment (LUS)] during LUS-muscular-thickness measurement, (B) to intro...
Gespeichert in:
Veröffentlicht in: | Archives of gynecology and obstetrics 2018-08, Vol.298 (2), p.297-306 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
The purpose of this study was: (A) to establish the effects of different ultrasound measurement methods (linear versus curved array) and measuring conditions [impact of pressure by fetal head/pelvis on the lower uterine segment (LUS)] during LUS-muscular-thickness measurement, (B) to introduce the intraoperative ultrasound measurement of LUS-muscular thickness (reference measurement method), and (C) to evaluate the correlation between different combinations of LUS-muscular-thickness measurement ultrasound techniques at birth planning and preoperative versus intraoperative measurements.
Methods
A prospective clinical observational study of women with the previous cesarean delivery was conducted. LUS-muscular thickness was measured: first at birth planning and second preoperatively using linear and curved probes (transabdominal) and an endocavitar probe (transvaginal), examined with and without pressure by fetal head/pelvis on LUS during measurement and third intraoperatively during repeat cesarean. Bland–Altman plots, paired
t
tests, Pearson’s correlation coefficient, and scatter plots were used.
Results
Thirty-three women were included in the study (ultrasound measurements:
n
= 601). There was no systematic difference between LUS-muscular-thickness measurements with linear versus curved array (mean difference = 0.06 mm;
p
= 0.24;
n
m
= 133) but between measurements with pressure by the fetus versus without (mean difference = − 0.37 mm;
p
|
---|---|
ISSN: | 0932-0067 1432-0711 |
DOI: | 10.1007/s00404-018-4805-6 |