Detector for positronium temperature measurements by two-photon angular correlation
We report on the design and characterization of a modular γ-ray detector assembly developed for accurate and efficient detection of coincident 511 keV back-to-back γ-rays following electron-positron annihilation. Each modular detector consists of 16 narrow lutetium yttrium oxyorthosilicate scintilla...
Gespeichert in:
Veröffentlicht in: | Review of scientific instruments 2018-05, Vol.89 (5), p.053106-053106 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report on the design and characterization of a modular γ-ray detector assembly developed for accurate and efficient detection of coincident 511 keV back-to-back γ-rays following electron-positron annihilation. Each modular detector consists of 16 narrow lutetium yttrium oxyorthosilicate scintillators coupled to a multi-anode Hamamatsu H12700B photomultiplier tube. We discuss the operation and optimization of 511 keV γ-ray detection resulting from testing various scintillators and detector arrangements concluding with an estimate of the coincident 511 keV detection efficiency for the intended experiment and a preliminary test representing one-quarter of the completed array. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.5017724 |