Evaluation and comparison of automated hematology analyzer, flow cytometry, and digital morphology analyzer for monocyte counting
Introduction This study was aimed to evaluate monocyte counts on Sysmex XN‐9000, Sysmex CyFlow Space System, and Sysmex DI60 and compare the performance of these systems with the reference optical microscopy (OM) assessment. Methods In all, 55 peripheral blood samples, collected in K3EDTA tubes, wer...
Gespeichert in:
Veröffentlicht in: | International journal of laboratory hematology 2018-10, Vol.40 (5), p.577-585 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
This study was aimed to evaluate monocyte counts on Sysmex XN‐9000, Sysmex CyFlow Space System, and Sysmex DI60 and compare the performance of these systems with the reference optical microscopy (OM) assessment.
Methods
In all, 55 peripheral blood samples, collected in K3EDTA tubes, were analyzed with XN‐9000, CyFlow System (FlowDiff1 and 2), DI60, and OM. Within‐run imprecision was carried out using normal samples. Data comparison was performed with Passing‐Bablok regression and Bland‐Altman plots.
Results
The within‐run imprecision of monocyte count on XN, FlowDiff, OM, and DI60 ranged between 1.9% for FlowDiff 2 and 22.1% for DI60. The Passing‐Bablok regression analysis of absolute count yielded slopes comprised between 0.93 (FlowDiff2 vs DI60) and 1.21 (DI60 vs OM), whereas the intercepts ranged between −0.002 (FlowDiff 1 vs FlowDiff 2) and 0.13 (FlowDiff1 and 2 vs DI60). Bland‐Altman plots in absolute values yielded absolute bias comprised between −0.01 × 109/L (FlowDiff 1 vs FlowDiff 2; DI60 vs OM) and 0.15 × 109 (XN‐module vs DI60).
Conclusion
The results of this analytical evaluation suggest that flow cytometry generates monocyte counts suitable for routine clinical use. OM or DI60 analysis may be useful for identifying morphologic abnormalities, but does not achieve a satisfactory level of accuracy for enumerating blood cells types such as monocytes, which are usually very low in peripheral blood. |
---|---|