EXPERIMENTAL AND COMPUTATIONAL EVALUATION OF EFFECTIVE CENTRE FROM A LONG COUNTER AT NEUTRON METROLOGY LABORATORY IN BRAZIL
A long counter detector was manufactured by the Institute of Advanced Studies (IEAV) and was characterised in the neutron low scattering room at Brazilian National Ionising Radiation Metrology Laboratory (LNMRI/IRD) to deploy a secondary Standard for neutron fluence. The effective centre was measure...
Gespeichert in:
Veröffentlicht in: | Radiation protection dosimetry 2018-08, Vol.180 (1-4), p.37-41 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A long counter detector was manufactured by the Institute of Advanced Studies (IEAV) and was characterised in the neutron low scattering room at Brazilian National Ionising Radiation Metrology Laboratory (LNMRI/IRD) to deploy a secondary Standard for neutron fluence. The effective centre was measured experimentally with 252Cf+D2O, 252Cf, 241AmBe and 238PuBe neutron sources, having average energies from 0.55 to 4.16 MeV. The experimental arrangement and detector construction were carefully reproduced in Monte Carlo simulations, and the computational results were found to be in good agreement with those from experiment. |
---|---|