The best of two worlds? Pulmonary vein isolation using a novel radiofrequency ablation catheter incorporating contact force sensing technology and 56-hole porous tip irrigation
Aims This study aimed to evaluate feasibility and safety as well as 1-year clinical outcome of pulmonary vein isolation (PVI) using a unique radiofrequency ablation catheter (“Thermocool SmartTouch SurroundFlow”; STSF) incorporating both, contact force (CF) sensing technology and enhanced tip irriga...
Gespeichert in:
Veröffentlicht in: | Clinical research in cardiology 2018-11, Vol.107 (11), p.1003-1012 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aims
This study aimed to evaluate feasibility and safety as well as 1-year clinical outcome of pulmonary vein isolation (PVI) using a unique radiofrequency ablation catheter (“Thermocool SmartTouch SurroundFlow”; STSF) incorporating both, contact force (CF) sensing technology and enhanced tip irrigation with 56 holes, in one device.
Methods
A total of 110 patients suffering from drug-refractory atrial fibrillation underwent wide area circumferential PVI using either the STSF ablation catheter (75 consecutive patients, study group) or a CF catheter with conventional tip irrigation (“Thermocool SmartTouch”, 35 consecutive patients, control group). For each ablation lesion, a target CF of ≥ 10–39 g and a force time integral (FTI) of > 400 g s was targeted.
Results
Acute PVI was achieved in all patients with target CF obtained in > 85% of ablation points when using either device. Mean procedure time (131.3 ± 33.7 min in the study group vs. 133.0 ± 42.0 min in the control group;
p
= 0.99), mean fluoroscopy time (14.0 ± 6 vs. 13.5 ± 6.6 min;
p
= 0.56) and total ablation time were not significantly different (1751.0 ± 394.0 vs. 1604.6 ± 287.8 s;
p
= 0.2). However, there was a marked reduction in total irrigation fluid delivery by 51.7% (265.52 ± 64.4 vs. 539.6 ± 118.2 ml;