Stereotactically navigated percutaneous microwave ablation (MWA) compared to conventional MWA: a matched pair analysis

Purpose To compare CT-navigated stereotactic microwave ablation (SMWA) to non-navigated conventional MWA (CMWA) for percutaneous ablation of liver malignancies. Methods A matched pair analysis of 36 patients who underwent MWA of primary or secondary liver malignancies (10 hepatocellular carcinoma an...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal for computer assisted radiology and surgery 2018-12, Vol.13 (12), p.1991-1997
Hauptverfasser: Beyer, L. P., Lürken, L., Verloh, N., Haimerl, M., Michalik, K., Schaible, J., Stroszczynski, C., Wiggermann, P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose To compare CT-navigated stereotactic microwave ablation (SMWA) to non-navigated conventional MWA (CMWA) for percutaneous ablation of liver malignancies. Methods A matched pair analysis of 36 patients who underwent MWA of primary or secondary liver malignancies (10 hepatocellular carcinoma and 8 metastases) was conducted. A total of 18 patients undergoing SMWA were included in this prospective study. Patients were matched in terms of tumor size, liver segment and entity to retrospective CMWA procedures. The endpoints were procedure time, accuracy of needle placement, technical success rate, complication rate and dose-length product (DLP). Results The procedure durations were 23.9 min (SD 3.7) for CMWA and 21.8 min (SD 16.3) for SMWA ( p = 0.22). The procedural accuracy for SMWA and CMWA was identical for both groups (3.7 mm). The total DLP was significantly lower for SMWA than for CMWA (2115 mGy cm (SD 276) vs. 3109 mGy cm (SD 1137), respectively; p  
ISSN:1861-6410
1861-6429
DOI:10.1007/s11548-018-1778-7