Assessment of the renal collecting system using a pocket-sized ultrasound device

Purpose To assess the performance of a pocket-sized ultrasound device (PUD) for evaluating dilatation of the renal collecting system with high-end ultrasound devices (HUDs) as a reference standard. Methods One sonographer examined both kidneys using a PUD to evaluate dilatation of the collecting sys...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of medical ultrasonics (2001) 2018-10, Vol.45 (4), p.577-581
Hauptverfasser: Kameda, Toru, Uebayashi, Kumiko, Wagai, Kazuko, Kawai, Fukiko, Taniguchi, Nobuyuki
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose To assess the performance of a pocket-sized ultrasound device (PUD) for evaluating dilatation of the renal collecting system with high-end ultrasound devices (HUDs) as a reference standard. Methods One sonographer examined both kidneys using a PUD to evaluate dilatation of the collecting system. The grading of the dilatation ranged from 0 to 4. Immediately after the examination, another sonographer blinded to the previous results performed a formal examination with a HUD. Results Two hundred kidneys in 100 patients were included in the analysis. The agreement of grades between the PUD and HUDs was excellent (weighted kappa = 0.83; P  
ISSN:1346-4523
1613-2254
DOI:10.1007/s10396-018-0881-2