Accuracy of iodine quantification in dual-layer spectral CT: Influence of iterative reconstruction, patient habitus and tube parameters
Evaluation of the influence of iterative reconstruction, tube settings and patient habitus on the accuracy of iodine quantification with dual-layer spectral CT (DL-CT). A CT abdomen phantom with different extension rings and four iodine inserts (1, 2, 5 and 10 mg/ml) was scanned on a DL-CT. The phan...
Gespeichert in:
Veröffentlicht in: | European journal of radiology 2018-05, Vol.102, p.83-88 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Evaluation of the influence of iterative reconstruction, tube settings and patient habitus on the accuracy of iodine quantification with dual-layer spectral CT (DL-CT).
A CT abdomen phantom with different extension rings and four iodine inserts (1, 2, 5 and 10 mg/ml) was scanned on a DL-CT. The phantom was scanned with tube-voltages of 120 and 140 kVp and CTDIvol of 2.5, 5, 10 and 20 mGy. Reconstructions were performed for eight levels of iterative reconstruction (i0-i7). Diagnostic dose levels are classified depending on patient-size and radiation dose.
Measurements of iodine concentration showed accurate and reliable results. Taking all CTDIvol-levels into account, the mean absolute percentage difference (MAPD) showed less accuracy for low CTDIvol-levels (2.5 mGy: 34.72%) than for high CTDIvol-levels (20 mGy: 5.89%). At diagnostic dose levels, accurate quantification of iodine was possible (MAPD 3.38%). Level of iterative reconstruction did not significantly influence iodine measurements. Iodine quantification worked more accurately at a tube voltage of 140 kVp. Phantom size had a considerable effect only at low-dose-levels; at diagnostic dose levels the effect of phantom size decreased (MAPD |
---|---|
ISSN: | 0720-048X 1872-7727 |
DOI: | 10.1016/j.ejrad.2018.03.009 |