Evaluation of image quality and radiation dose of thoracic and coronary dual-source CT in 110 infants with congenital heart disease
Background: There are only a few reports on the diagnostic accuracy, and the technical and clinical feasibility, of multidetector CT (MDCT) in infants with congenital heart disease (CHD). Objective: To evaluate the image quality and radiation dose of DSCT in babies with CHD. Materials and methods: F...
Gespeichert in:
Veröffentlicht in: | Pediatric radiology 2009-07, Vol.39 (7), p.668-676 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background:
There are only a few reports on the diagnostic accuracy, and the technical and clinical feasibility, of multidetector CT (MDCT) in infants with congenital heart disease (CHD).
Objective:
To evaluate the image quality and radiation dose of DSCT in babies with CHD.
Materials and methods:
From November 2006 to November 2007, 110 consecutive infants with CHD referred for pre- or postoperative CT evaluation were included. All these infants had a spiral angiothoracic DSCT scan after injection of 300 mg/ml iopromide at 0.5–1 ml/s with a power injector using a low-dose protocol (80 kVp and 10 mAs/kg). Of these infants, 34 also underwent an ECG-gated coronary CT scan for evaluation of the course of the coronary arteries.
Results:
No serious adverse events were recorded. The mean dose-length product was 8±6 mGy.cm (effective dose 0.5±0.2 mSv) and 21±9 mGy.cm (effective dose 1.3±0.6 mSv) during the non-ECG-gated spiral acquisition and ECG-gated acquisition, respectively. Diagnostic quality images were achieved with the spiral acquisition in 89% of cases. Compared to the spiral mode, ECG-gated acquisition significantly improved the visualization of the coronary arteries, with a diagnostic rate of 91% and 84% for the left and right coronary arteries, respectively.
Conclusion:
DSCT together with iopromide at 300 mg/ml is a valuable tool for the routine clinical evaluation of infants with CHD. ECG-gated acquisition provides reliable visualization of the course of the coronary arteries. |
---|---|